To adjust the start pressure of the ramp on your Z1:
- With the Z1 is in standby mode, press and hold the Down Arrow Button for approximately three seconds until the 4.0 (or another pressure reading) begins to flash.
- While the start pressure value is flashing, repeatedly press either of the Up/Down buttons to increase or decrease the start pressure.
- When the desired ramp start pressure is reached, press and hold the Start/Stop Button for approximately three seconds to save your setting (this setting will be saved even when the Z1 is not powered). The Z1 will return to standby mode.
NOTE: To return to standby mode without saving the new setting, press and release the start/Stop button momentarily instead of holding it down.
You can also easily change the ramp pressure of your Z1 Auto via the Nitelog app.